The didymium liquid filter consists of praseodymium and neodymium, dissolved in perchloric acid. This filter is ideally suited for checking the wavelength accuracy of spectrophotometers in the UV and visible range. It has a spectrum with a variety of characteristic, very well-defined peaks in the range between 320 nm and 870 nm.
This filter is recommended in the Ph. Eur. and in the USP 857 for checking the wavelength accuracy above 640 nm.
Ph. Eur. = European Pharmacopoeia
USP = United States Pharmacopeia
Typical spectrum of Didymium (praseodymium and neodymium) dissolved in perchlroic acid measured at a slit width of 1 nm.
ARTICLE-NO. | 667025 Request this article |
APPLICATION | Checking the wavelength accuracy in the UV and Vis range |
CONTENT | Didymium in perchloric acid |
STANDARD MEASUREMENT | Wavelengths: 329, 469, 575, 740, 864 nm Slit width: 1 nm; with DAkkS calibration certificate |
POSSIBLE MEASUREMENTS | Wavelengths: 329, 354, 444, 469, 482, 512, 522, 575, 732, 740, 794, 801, 864 nm Slit width: 1 - 2 nm |
ARTICLE-NO. | 667025EPUSP Request this article |
APPLICATION | Checking the wavelength accuracy according to Ph. Eur. and USP <857> in the Vis range above 640 nm |
CONTENT | Didymium in perchloric acid |
STANDARD MEASUREMENT | Wavelengths: 512, 732, 740, 794, 801, 864 nm Slit width: 1 nm; with DAkkS calibration certificate |
POSSIBLE MEASUREMENTS | Slit width: 1 - 2 nm |