Article-No. | Application | Content | Certified wavelength in nm |
---|---|---|---|
667EP1001 | Checking the photometric accuracy in UV-range | UV506, 6 mg/l Niacin in HCl, (0.25 Abs) UV512, 12 mg/l Niacin in HCl, (0.5 Abs) UV524, 24 mg/l Niacin in HCl, (1.0 Abs) UV599, Hydrochloric acid (HCl), reference filter | 213, 261 |
Checking the photometric accuracy in the Vis-range | Neutral Density Glass Filters: F2 (0.25 Abs) F3 (0.5 Abs F4 (1.0 Abs) | 440, 465, 546.1, 590, 635 | |
Checking the wavelength accuracy | UV5, Holmium oxide in HClO4 UV14, Perchloric acid, reference filter | 241, 287, 361, 451, 485, 536, 640 | |
Checking stray light | UV11, Sodium Nitrite in H2O UV12, Pure Water (H2O), reference filter Acsorbance > 3.0 Abs at 340 nm and 370 nm | Cut-off: 388 | |
Checking the spectral resolution | UV6, Toluene in Hexane UV9, Hexane, reference filter | Scan: 265 - 270 Slit widths: 0.5, 1.0, 1.5, 2.0, 3.0 |
Article-No. | Application | Content | Certified wavelength in nm |
---|---|---|---|
667EP1002 | Checking the photometric accuracy in UV-range | UV506, 6 mg/l Niacin in HCl, (0.25 Abs) UV512, 12 mg/l Niacin in HCl, (0.5 Abs) UV524, 24 mg/l Niacin in HCl, (1.0 Abs) UV599, Hydrochloric acid (HCl), reference filter | 213, 261 |
Checking the wavelength accuracy | UV5, Holmium oxide in HClO4 UV14, Perchloric acid, reference filter | 241, 287, 361, 451, 485, 536, 640 | |
Checking stray light | UV23, Potassium Iodide in H2O UV12, Pure Water (H2O), reference filter Acsorbance > 3.0 Abs at 250 nm | Cut-off: 259 | |
Checking the spectral resolution | UV6, Toluene in Hexane UV9, Hexane, reference filter | Scan: 265 - 270 Slit widths: 0.5, 1.0, 1.5, 2.0, 3.0 |
Article-No. | Application | Content | certified wavelength in nm |
---|---|---|---|
667857V3 | Checking the photometric accuracy and linearity in the UV range | Potassium Dichromate Filters: UV20, 20 mg/l (0.1 - 0.3 Abs) UV60, 60 mg/l (0.3 - 0.9 Abs) UV0140, 140 mg/l (0.7 - 2.0 Abs) UV14, Percloric acid (reference filter) | 235, 257, 313, 350 |
Checking the photometric accuracy and linearity in the Vis range | Neutral Density Glass Filters: F2 (0.25 Abs) F4 (1.0 Abs) F203 (2.0 Abs) | 440, 465, 546.1, 590, 635 | |
Checking the wavelength accuracy | UV5, Holmium oxide in Perchloric acid | UV5: 241, 250, 278, 287, 333, 345, 361, 386, 416, 451, 468, 485, 536, 640 | |
Checking stray light | UV19, Acetone (pure), ≥ 2 Abs at 300 nm, measured against air | Spectral range: 250 - 330 | |
Checking the spectral resolution | UV6, Toluene in Hexane UV9, Hexane, reference filter | Scan: 265 - 270 Slit widths: 0.5, 1.0, 1.5, 2.0, 3.0 |
Article-No. | Application | Content | certified wavelength in nm |
---|---|---|---|
667857V4 | Checking the photometric accuracy in the UV range | Potassium Dichromate Filters: UV20, 20 mg/l (0.1 - 0.3 Abs) UV0140, 140 mg/l (0.7 - 2.0 Abs) UV14, Percloric acid (reference filter) | 235, 257, 313, 350 |
Checking the photometric accuracy in the Vis range | Neutral Density Glass Filters: F2 (0.25 Abs) F203 (2.0 Abs) | 440, 465, 546.1, 590, 635 | |
Checking the wavelength accuracy | UV5, Holmium oxide in Perchloric acid | UV5: 241, 250, 278, 287, 333, 345, 361, 386, 416, 451, 468, 485, 536, 640 | |
Checking stray light | UV19, Acetone (pure), ≥ 2 Abs at 300 nm, measured against air | Spectral range: 250 - 330 | |
Checking the spectral resolution | UV6, Toluene in Hexane UV9, Hexane, reference filter | Scan: 265 - 270 Slit widths: 0.5, 1.0, 1.5, 2.0, 3.0 |