Rare Earth (Cerium) liquid filter (200- 260 nm)

The rare earth (Cerium) liquid filter consists of a solution of rare earth metals dissolved in perchloric acid. This filter is ideally suited for checking the wavelength accuracy of spectrophotometers in the low UV range. It has a spectrum with five characteristic peaks in the range from 200 nm to 260 nm.

This filter is recommended in the Ph. Eur. and in the USP 857 for checking the wavelength accuracy in the range of 200 - 260 nm.

Ph. Eur. = European Pharmacopoeia
USP = United States Pharmacopeia

Typical spectrum of a Rare Earth (Cerium) filter, measured at a slit width of 1 nm

ARTICLE-NO. 667035   Request this article
APPLICATION Checking wavelength accuray according to Ph. Eur. and USP <857> in the UV range below 240 nm
CONTENT Rare earth (Cerium) in perchloric acid
STANDARD MEASUREMENT Wavelengths: 201, 211, 222, 239, 252 nm
Slit width: 1 nm; with DAkkS calibration certificate

Optical Reference Materials (Brochure & User Manual EN)

Download

Contact/Inquiry

Hellma USA Inc.
120 Terminal Drive
11803 Plainview, NY
United States

Phone: +1 5169 3908 88
info.us@hellma.com
www.hellmausa.com

Contact/Inquiry

Disabled, please use email.